Product Introduction
MATS-3000S Hysteresisgraph can automatically measure the DC/AC (DC ~ 1kHz)magnetic parameters of soft magnetic materials, such as: saturation field Bs, remanence Br, coercivity Hc, total loss ratio Ps, amplitude permeability µa, incremental permeability µm, and so on.
Applied Standards
- IEC 60404-4 & IEC 60404-6 measurement methods of magnetic properties of magnetic soft materials
Technical Parameters
Under DC , technical data of measuring ring sample of metal magnetic power core:
Parameters |
Bs(%) |
Br(%) |
Hc(%) |
Pu(%) |
μm(%) |
μi(%) |
Uncertainty(k=2) |
1 |
1 |
2 |
1.5 |
2 |
5 |
Repeatability |
±0.5 |
±0.5 |
± 1 |
± 1 |
± 1 |
± 3 |
Remarks | Test by locking Hs, the locking accuracy 0.5%. |
Under AC frequency of 10Hz~1kHz, technical data of measuring amorphous ring sample are as follows:
Parameters |
Ps(%) |
μa(%) |
Bm(%) |
Br(%) |
Hc(%) |
|
Uncertainty(k=2) |
3 |
3 |
1 |
1 |
2 |
|
Repeatability |
± 2 |
± 1 |
± 0.5 |
± 0.5 |
± 1 |
|
Remarks | Test by locking Bm, the lock accuracy 0.5%. Test by locking Hm, the lock accuracy 0.5%. |
Under AC frequency of 0.01Hz~10Hz, technical data of measuring ferrite ring sample are as follows:
Parameters |
μa(%) |
Bm(%) |
Br(%) |
Hc(%) |
||
Uncertainty(k=2) |
3 |
2 |
2 |
4 |
||
Repeatability |
± 1 |
± 1 |
± 1 |
± 2 |
||
Remarks | Samples should be thin ring, OD/ID≈1.25. Test by lock Hm, the lock accuracy 0.5%. Demagnetizing before testing; the frequency of demagnetizing field ≤ test frequency. |
Notes:
- The desk in the picture is NOT included in the configuration.
- Optional laser printer is available.
- Components may appear different from the picture.